SR7000 and SR7000DC Exclusive Features
1. HPLC Fluidics
a. User Serviceable
b. nano to micro to macro applications
0.0025" to 0.04" I.D. Tubing
c. Allows flow of cell lysates
2. Sapphire Prism
a. Virtually indestructible
b. High thermal conductivity contributes to exceptional thermal stability
3. 10 Deg. C. to 90 Deg. C. temperature range
Temperature changes programmable in user defined steps and intervals
4. Flow or Static Sample Cell
a. Available micro and macro flow cells
5. Proprietary Image analysis software
Contributes to baseline stability
6. 3700 Pixel linear CCD array image sensor
Extremely high resolving power contributes to low peak to peak noise and baseline stability
7. Illumination Level Feedback Control
Separate photodiode light level detector and feedback circuitry
stabilizes laser LED
Improves peak to peak noise and baseline stability
8. Broad Refractive Index range - 1.32 to 1.6
Useful for Organic Solvents (DMSO), Urea
9. Full access to Reflectivity Data
High quality optics enables generation of useful reflectivity data
Reflectivity vs. angle data may be downloaded to computer and "Fit" to Fresnel
Models to determine layer thickness, refractive index
Source code is available to instrument purchasers for customization/modification
10. Sensor Slides
Planar Carboxyl surface in Polyethylene Glycol Background.
Extremely low non-specific binding surface
Proven high quality kinetic data without diffusion or ionic interference's
Also, Streptavidin, NiNTA, Polylysine and plain gold
|